Should Glucose Control Solution Register

What is the purpose of command solution testing?

Performing a glucose control exam is an excellent way to bank check that your system (meter and strips) are working optimally to deliver accurate and precise results.

Using manufacturer's command solution to test the system, the results volition provide information to determine if the organization passes or fails the test. If the system fails the test, farther data needs to exist obtained to troubleshoot the system.

What are the ingredients in control solution?

Command solution is a mixture of sugar, water and other stabilizers. Unlike a blood sample, command solution does not contain any blood cells, interferents, or blood products and is less viscous than blood. Control Solution necessarily reacts differently than blood (move viscous) when applied to the examination strip. Therefore, the results of a glucose control test exercise not reverberate a true blood glucose value. DO NOT Ever Drink OR INJECT THE Control SOLUTION!

How are control solutions ranges developed?

Manufacturers develop control solution(s) and assign acceptable ranges for each level of control solution (level 0, 1, two, etc.). The ranges are based on regulatory guidelines, the accuracy and precision of examination strips and glucose control solution.

What are the expiration dates for command solution?

Control_solution_expiration_dates

Control solution has two expiration dates, (1) is the closed bottle expiration date which is printed on the glucose control bottle characterization and (2) is 3 months from the appointment the command solution bottle was first opened – please refer to the control solution Instructions for Use for more information. It is practiced do to write the engagement when the solution was kickoff opened on the canteen label.

How frequently should I use the control solution to test my meter?

We recommend using Control solution to test the meter:

• Before using the arrangement to perform a blood test

• When opening a new vial of strips

• If blood results seem unusually high or low based on patient's symptoms and status

• If strip vial has been left open up or exposed to extreme heat, cold or humidity (recommend replacement)

• Whenever a cheque on the performance of the system is needed

• If meter damage is suspected (Meter dropped, crushed, moisture, etc.)

Should I shake the glucose control solution prior to using information technology?

No, control solution should be gently swirled or inverted two-three times earlier utilise. DO Non Shake! Shaking the vial of solution will introduce air bubbles into the solution, which may affect the results.

Does temperature bear upon the glucose control results?

The meter, examination strips and control solution should to exist at room temperature, (68-77◦F) for at least 30 minutes prior to command testing. Control solution and the test strips are afflicted by temperatures below 68◦ F and above 77◦F, as well every bit humidity. The solution can exist stored at temperatures 36◦-86◦, but brought to room temperature at to the lowest degree xxx minutes prior to testing.

How practise I perform a control test?

control_solution_tip

control solution

control solution with meter

To go an authentic reading, first clasp a drop of control solution onto a tissue, and so wipe the tip of the bottle clean. The solution is basically sugar in water with a few additives, and tin dry out in the bottle tip equally the water evaporates, leaving a concentrated solution on the tip, which may result in an inaccurate reading. The next step should be to apply a drop of control solution onto a clean not-porous surface, such as aluminum foil or plastic wrap and affect the tip of the test strip into the solution. Remove the test strip from the drop of solution as soon as the meter beeps and begins testing.

How practise I interpret the results of the command solution exam?

The command ranges are located on the vial of test strips that you are using.

Choose the upshot range matching the number on your glucose command solution vial to meet if your system has passed or failed the glucose control exam. If the result on the meter is within the range on the vial that matches the number on the control solution, the meter, strips and solution pass.

If the result on the meter is non inside the acceptable range, then additional troubleshooting needs to occur.

The control solution test effect was at the pinnacle limit (or low limit) of the acceptable range printed on the vial. Does this mean my blood glucose examination results volition be on the loftier terminate (or low end) also?

No, the control solution event does not correlate with claret glucose results. Control testing must exist interpreted as Pass/Fail. Where the command exam results autumn within the indicated test-range is in no way an indication of blood test results. Control solution has a completely unlike chemical and concrete makeup compared to blood. These differences provide for stability and increased sensitivity to many conditions that could affect the blood glucose exam arrangement. Control solution'south increased stability and sensitivity make it ideal for field assessment of the glucose measurement system. If the result is within the range printed on the test strip vial label, then the glucose meter, strips and solution all pass the testing and are working properly.

Diabetic Warehouse is a trusted supplier of diabetes intendance products and accessories. For more information and to explore a consummate range of products, including test strips, syringes and needles, glucose monitoring systems, and more, visit www.diabeticwarehouse.org .

Should Glucose Control Solution Register,

Source: https://www.diabeticwarehouse.org/blogs/articles/frequently-asked-questions-regarding-glucose-control-solution-testing

Posted by: caballerosholl1946.blogspot.com

0 Response to "Should Glucose Control Solution Register"

Post a Comment

Iklan Atas Artikel

Iklan Tengah Artikel 1

Iklan Tengah Artikel 2

Iklan Bawah Artikel